Research Article

Guidewire Ablation within the Coronary Venous System for Epicardial or Intramural Ventricular Arrhythmia: A Preclinical Study of Biophysical Characterization

Table 4

Orthogonal design and experimental results.

No.Power (w)Duration (s)Infusion speed (ml/min)Maximum diameter (mm)Steam pop incidence

1105013.600/8
2106023.200/8
3107032.100/8
4155024.220/8
5156033.100/8
6157015.253/8
7205033.410/8
8206015.404/8
9207025.131/8
K12.963.744.75ā€”
K24.193.894.17
K34.654.162.87
R1.690.421.88