Research Article

A Simple and Rapid LC-MS/MS Method for the Quantification of Nirmatrelvir/Ritonavir in Plasma of Patients with COVID-19

Table 3

Inter- and intraday precision and accuracy of analytes (n= 5).

AnalytesNominal concentration (ng/mL)Interday measured concentration (ng/mL) ± SDRSD (%)RE (%)Intraday measured concentration (ng/mL) ± SDRSD (%)RE (%)

Nirmatrelvir54.8 ± 0.47.9−3.95.0 ± 2.29.90.1
1000983.3 ± 89.19.1−1.7982.3 ± 78.07.9−1.8
20002050 ± 105.65.22.52088 ± 104.45.04.4

Ritonavir54.8 ± 0.713.64.15.1 ± 0.713.41.5
10001031 ± 118.711.53.11078 ± 18.11.77.8