Research Article

A Simple and Rapid LC-MS/MS Method for the Quantification of Nirmatrelvir/Ritonavir in Plasma of Patients with COVID-19

Table 5

Results of stability in the autosampler of analytes (n = 3).

AnalytesNominal concentration (ng/mL)0 h12 h24 h
Mean (%) ± SDRSD (%)RE (%)Mean (%) ± SDRSD (%)RE (%)Mean (%) ± SDRSD (%)RE (%)

Nirmatrelvir55.1 ± 0.24.52.04.38 ± 0.214.8−12.55.4 ± 0.34.97.2
10001077 ± 32.43.07.7908.4 ± 65.57.2−9.2982.2 ± 29.93.0−1.8
20002140 ± 78.13.77.02004 ± 15.60.80.22054 ± 71.33.52.7

Ritonavir54.6 ± 0.23.6−8.04.51 ± 0.163.4−9.84.5 ± 0.23.7−9.1
10001110 ± 25.92.311.01079 ± 73.16.87.9897.5 ± 65.27.310.3
20002120 ± 33.11.66.01918 ± 12.20.64.11940 ± 46.82.4−3.0