Research Article
Analytical and Clinical Evaluation of a Chemiluminescent Immunoassay to Detect Serum Chitinase-3-like Protein 1 in HBV-Related Liver Diseases
Table 1
Analysis of the CHI3L1 linear range.
| Dilution ratio (L : H) | Expected concentration (ng/mL) | Measured value (ng/mL) | Detection average (ng/mL) | 1st | 2nd |
| 5 : 0 | 1.53 | 1.52 | 1.54 | 1.53 | 4 : 1 | 398.42 | 395.42 | 395.41 | 395.42 | 3 : 2 | 795.32 | 775.26 | 775.41 | 775.34 | 2 : 3 | 1192.21 | 1145.00 | 1125.00 | 1135.00 | 1 : 4 | 1589.11 | 1524.00 | 1527.00 | 1525.50 | 0 : 5 | 1986.00 | 1985.00 | 1987.00 | 1986.00 |
|
|