Research Article

Analytical and Clinical Evaluation of a Chemiluminescent Immunoassay to Detect Serum Chitinase-3-like Protein 1 in HBV-Related Liver Diseases

Table 1

Analysis of the CHI3L1 linear range.

Dilution ratio (L : H)Expected concentration (ng/mL)Measured value (ng/mL)Detection average (ng/mL)
1st2nd

5 : 01.531.521.541.53
4 : 1398.42395.42395.41395.42
3 : 2795.32775.26775.41775.34
2 : 31192.211145.001125.001135.00
1 : 41589.111524.001527.001525.50
0 : 51986.001985.001987.001986.00

L, low; H, high; CHI3L1, chitinase-3-like protein 1.