Research Article

Analytical and Clinical Evaluation of a Chemiluminescent Immunoassay to Detect Serum Chitinase-3-like Protein 1 in HBV-Related Liver Diseases

Table 4

Evaluation of precision.

LevelMean (ng/mL)Intraassay precisionInterassay precisionTotal precision
SDCV (%)SDCV (%)SDCV (%)

Level 119.730.361.811.537.761.567.91
Level 285.272.352.761.371.612.462.88

SD, standard deviation; CV, coefficient of variation. CV (%) = (SD/mean) × 100%.