Fluke 115 AC/DC digital multimeter technical and general specifications [35].
(a) Accuracy specifications

FunctionRangeResolutionAccuracy ± ([% of reading] + [counts])

Millivolts600.0 mV0.1 mV0.5% + 2

DC volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5% + 2

AC millivolts
true-rms
600.0 mV0.1 mV45 Hz to 500 Hz: 1.0% + 3
500 Hz to 1 kHz: 2.0% + 3

AC volts
true-rms
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
45 Hz to 500 Hz: 1.0% + 3
500 Hz to 1 kHz: 2.0% + 3

Continuity600 Ω1 ΩDetects opens or shorts of 500 µs or longer

Ohms600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.9% + 2
0.9% + 1
0.9% + 1
0.9% + 1
0.9% + 1
1.5% + 2

Diode test2.000 V0.001 V0.9% + 2

Capacitance1000 nF
10.00 µF
100.0 µF
9999 µF
1 nF
0.01 µF
0.1 µF
1 µF
1.9% + 2
1.9% + 2
1.9% + 2
100 µF to 1000 µF: 1.9% + 2
>1000 µF: 5% + 20

AC amps (45 Hz to 500 Hz)6.000 A
10.00 A
20 A overload for 30 seconds maximum
0.001 A
0.01 A
1.5% + 3

DC amps6.000 A
10.00 A
20 A overload for 30 seconds maximum
0.001 A
0.01 A
1.0% + 3

Hz (V or A input)99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1% + 2

(b) Frequency counter sensitivity

Input rangeTypical sensitivity (rms sine wave)
5 Hz to 45 Hz45 Hz to 5 kHz5 kHz to 10 kHz10 kHz to 50 kH

Volts AC6 V
60 V
600 V
0.2 V
2 V
20 V
0.2 V to 0.3 V
2 V to 3 V
0 V to 30 V
0.3 V to 0.4 V
3 V to 4 V
30 V to 40 V

AC amps6 A
10 A
N/A0.4 A
0.5 A
N/A

(c) Input characteristics

FunctionInput impedance (nominal)Common mode rejection ratio (1 kW unbalanced)Normal mode rejection

Volt AC>5 MΩ < 100 pF>60 dB at dc, 50 or 60 Hz

Volts DC>10 MΩ < 100 pF>100 dB at dc, 50 or 60 Hz>60 dB at 50 or 60 Hz

Auto-V LoZ~3 kΩ < 500 pF>60 dB at dc, 50 or 60 Hz

OhmsOpen-circuit test voltageFull scale voltageShort-circuit current
To 6.0 MΩTo 6.0 MΩ
<2.7 V dc<0.7 V dc<0.9 V dc<350 µA

Diode test2.7 V dc2.000 V dc<1.2 mA

(d) General specifications

FunctionDescription

Maximum voltage between any terminal and earth ground600 V

Surge protection6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2

W Fuse for A input11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)

DisplayDigital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec

TemperatureOperating: −10°C to +50°C; Storage: −40°C to +60°C

Humidity0% to 90% to 35°C
75% to 40°C
45% to 50°C

Temperature coefficient0.1 × (specified accuracy/°C) (<18°C or >28°C)

Operating altitude2,000 meters

Battery9 Volt Alkaline, NEDA 1604 A/IEC 6LR61

Battery lifeAlkaline: 400 hours typical, without backlight

Safety compliancesANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1

CertificationsUL, CSA, TUV, N10140, VDE

ModelFluke 115