Research Article

Multiple PAHs’ Detection under CDOM Interference Based on Excitation-Emission Matrix and Interval Selection

Table 4

Results of the recovery tests for four PAH measurements.

Spiked PAHsOriginal concentrationAdded concentrationMeasured concentrationRecovery (%)Average recovery (%)

FLU9.88 ± 0.192.3812.39 ± 0.31105.4103.3
9.88 ± 0.191.1911.17 ± 0.22108.4
9.88 ± 0.190.7110.57 ± 0.1797.2

PYR10.16 ± 0.232.3812.75 ± 0.20109.0103.2
10.16 ± 0.231.1911.32 ± 0.1997.7
10.16 ± 0.230.7110.89 ± 0.32102.8

PHE9.84 ± 0.392.3812.39 ± 0.14107.2107.4
9.84 ± 0.391.1911.14 ± 0.21109.5
9.84 ± 0.390.7110.59 ± 0.23105.6

BaP9.96 ± 0.142.3812.30 ± 0.1898.2101.4
9.96 ± 0.141.1911.19 ± 0.17103.5
9.96 ± 0.140.7110.69 ± 0.12102.4