Implementation of Bioelectrical Impedance Measuring Instrument Based on Embedded System

Table 3

Difference error between the bioelectrical impedance of human left and right legs at multi frequencies using the developed and standard device.

Frequency (kHz)Impedance (Ω)
Mean value
Left legRight leg
Standard deviceBIA instrumentDiff./error (%)Standard deviceBIA instrumentDiff./error (%)

5322.1311.13.42317.5302.34.79
50305.3291.24.62299.7283.65.37
100299.1286.84.11273.4280.22.49
150298.1282.15.37292.4275.45.81
200288.9276.44.33280.3269.63.82
250280.2272.12.89274.2264.23.65
300276.6270.62.17269.4262.42.60
400271.8267.31.66262.9260.70.84