Noninvasive Techniques for Blood Pressure Measurement Are Not a Reliable Alternative to Direct Measurement: A Randomized Crossover Trial in ICU

Table 1

Differences between patient systolic, diastolic, and mean arterial pressure taken with direct-invasive (IBP) and, respectively, with auscultatory-aneroid (ABP) and oscillometric automated (OBP) methods.

SystolicDiastolicMean

IBP
 Mean ± SD134.4 ± 24.157.9 ± 12.381.5 ± 14.5
 (median)(134.5)(57.0)(79.5)
ABP
 Mean ± SD124.7 ± 18.363.1 ± 12.983.7 ± 12.2
 (median)(125.0)(62.0)(82.0)
Difference (IBP − ABP)
 Mean ± SD9.7 ± 13.8−5.1 ± 7.1−2.1 ± 7.1
 (95% CI)(7.0; 12.5)(−6.5; −3.7)(−3.6; −0.7)
value

IBP
 Mean ± SD134.2 ± 24.358.0 ± 12.781.6 ± 14.6
 (median)(134.0)(57.0)(80.0)
OBP
 Mean ± SD123.4 ± 16.761.6 ± 11.976.2 ± 11.2
 (median)(125.0)(60.5)(75.5)
Difference (IBP − OBP)
 Mean ± SD10.8 ± 14.9−3.6 ± 6.05.4 ± 7.1
 (95% CI)(7.8; 13.8)(−4.8; −2.4)(4.0; 6.8)
value